Cassette Autoclave, STC-2000

คุณสมบัติสินค้า:

SKU : STC-2000

แบรนด์ : Infitek (China)

Share

Cassette Autoclave, STC-2000

  1. Sterilization Volume: 1.8L
  2. The whole sterilization process is about 6 minutes
  3. Temperature range: 135 ° C- 138° C
  4. Maximum Operating Pressure: 242 kPa

 

Description

  • The sterilization effect is reliable, rapid. The whole process of non-wrapped sterilization time is 6 minutes, suitable for stomatology, ophthalmology and other departments of small instruments.
  • Equipment size: 57 CM*41.5 CM*17 CM; sterilization chamber size: 28 CM*18 CM* 3.8 cm;
  • Sterilization box size (outside) : 39 CM *19.6 CM *4 CM.
    It is easy to operate, the system operation is displayed on the screen; Sterilization procedures can be safely interrupted at any time.
  • With self-detection system, if a problem occurs in the process of operation, the system will automatically stop and the defective code will be displayed on the screen.
  • It is suitable for sterilization of instruments containing cavity and solid instruments, and it can be non-wrapped sterilization.

 

  • Easy to operate
  • The system operation is displayed on the screen
  • Sterilization procedures can be safely interrupted at any time
  • Standard printer
  • Real-time record for sterilization process
  • Ultra-thin cassette box design makes heating and cooling faster
  • 0.22μm Biofilter

Shorter cycle time, higher efficiency

  • A. Air removal
  • B. Warming up
  • C. Pressurizing
  • D. Sterilizing
  • E. Venting

 

 

SPECIFICATIONS

STC-2000
Maximum Operating Pressure242 kPaInput Power≤1.3 kVA
Operating Pressure Range42 kPa ~ 212kPaAtmospheric Pressure Range70.0 kPa to 106.0 kPa
Maximum Operating Temperature138° CAmbient Temperature Range+ 5 ° C ~ + 40 ° C
Operating Temperature Range115 ° C – 135 ° CSteam Generator Power1.2 kVA ± 5%
Display accuracy0.1℃FuseF8AL250V phi 5 x 20 mm
Sterilization VolumeAbout 1.8 LitersSafety Valve Setting Pressure0.25 MPa
Tank Volume3.4 LitersEquipment Working MediumWater Vapor
WaterUltra-pure water, deionized water, laboratory water.Working Relative Humidity≤85%
Power Supply VoltageAC 220 V±22 VNormal Use PlaceWork room
Power Frequency50 Hz ± 1 HzRun ModeRun Intermittently (≤6 cycles/hour)


https://infitek.com/products/cassette-autoclave-stc-2000/

เว็บไซต์นี้มีการใช้งานคุกกี้ เพื่อเพิ่มประสิทธิภาพและประสบการณ์ที่ดีในการใช้งานเว็บไซต์ของท่าน ท่านสามารถอ่านรายละเอียดเพิ่มเติมได้ที่ นโยบายความเป็นส่วนตัว  และ  นโยบายคุกกี้