Products
Protein Biology Reagents | Consumables
Electrophoresis Gel Preparation
Electrophoresis Buffer
SKU :
Quantity
หมวดหมู่ : Biochemicals ,  1. Chemical and Reagents ,  Buffers ,  Servicebio , 
แบรนด์ : Servicebio
Share
Product Information
Product Name | Cat. No. | Spec. |
SWE Rapid High Resolution Running Buffer (Powder) | G2081-1L | 1 L |
G2081-15 | 1 L, 15 bags |
Product Description/Introduction
The SWE Rapid High Resolution Running Bufferr (powder) is optimal to Tris-Glycine electrophoresis buffer for SDS-PAGE gel electrophoresis in Tris-Glycine. Compared with the traditional Tris-glycine running buffer, this product can be used for electrophoresis without adjusting the voltage according to the separation gel and the stacking gel, and it can be used for the whole process of electrophoresis at a constant voltage of 200-250 V. It takes about 30 min to complete the electrophoresis, which saves time and is more effective for the separation of small molecules. It can separate the 10 kDa and the 15 kDa bands in the prestained protein marker and clearly visible (gel concentration above 8%) to achieve rapid and efficient isolation of
proteins.
Storage and Shipping Conditions
Ship and store at room temperature. Valid for 24 months.
Assay Protocol/Procedures
1. Each bag of electrophoresis buffer (powder) is dissolved in 1000 mL of purified water, and can be used after fully dissolved.
2. Selection of different protein gel concentrations for different protein molecular weights, refer to the table below:
SDS-PAGE Separation Gel Concentration | Optimum Separation Range (kDa) (SWE Rapid High Resolution Running Buffer) |
6% | |
8% | 10-250 |
10% | 5-150 |
12% | 3-100 |
15% | <60 |
3. Constant voltage electrophoresis, 200-250 V, Refer to the following table for electrophoresis time:
Comparison Table of Electrophoresis Voltage and Electrophoresis Time | |
Voltage(V) | Estimated Electrophoresis Time(min) |
200 | 35 |
220 | 30 |
250 | 25 |
Note
4. It is recommended to use the working solution as soon as possible after preparation.
5. This product is suitable for conventional Tris-HCl system SDS-PAGE gels, such as our products G2003, G2037, G2041, G2042,G2043,G2044,G2045,G2060,G2061,G2062, G2063,G2064.
6. High voltage electrophoresis is easy to produce high heat. Please select the appropriate voltage for electrophoresis according to the ambient temperature.
7. The running solution can be reused for 2-3 times, more reuse is not recommended.
8. For your safty and health, please wear safety glasses, gloves, or protective clothing.
For Research Use Only!