Auto Decalcification System

คุณสมบัติสินค้า:

SKU : SDH-5/SDH10

หมวดหมู่ : 12. Lab Equipment Pathatology Servicebio

แบรนด์ : Servicebio

Share

Auto Decalcification System
Provides a 4°C low-temperature environment to effectively protect the enzyme and protein activity in the sample. Supports decalcification of samples such as bone and teeth with a fast decalcification speed; a variety of specifications are available to reduce the amount of decalcification solution while ensuring high-quality sample processing.

Product Introduction
The Servicebio Decalcification System maintains samples in a 4°C low-temperature environment, effectively preserving enzyme and protein activity within the samples. In osteoblast staining, its performance is significantly superior to traditional methods, particularly for ALP staining, where it delivers clear and distinct results. Conventional decalcified samples often fail to achieve effective ALP staining due to enzyme inactivation.
Enzyme activity is highly sensitive to temperature and decalcification conditions. The Servicebio Decalcification System ensures precise preservation of enzyme and protein activity through continuous low-temperature control and optimized decalcification protocols, combined with specialized decalcification solution. This enables clear and effective staining results for osteoblast ALP staining.
Decalcification System(SDH-10/SDH-5) + EDTA Decalcification Solution (Slow Decalcification) (G1105-500ML) + Adhesive Slides(Bone Tissue-Specific) (G6078-50) + Repair Instrument—effortlessly resolve challenges posed by bone tissue samples.

Product Features

  • 4°C Low-Temperature Decalcification – Suitable for various samples, including bone tissue and teeth, ensuring fast processing and superior staining results.
  • Equipped with Adapters – Compatible with multiple sizes of centrifuge tubes, allowing direct sample and decalcification solution loading while minimizing reagent consumption.
  • High-Brightness Display – Ensures clear visibility of settings and progress.
  • Dual Temperature Display – Shows both the set temperature and real-time temperature simultaneously.
  • Dual Time Display – Displays both the programmed time and elapsed runtime.
  • Automatic Shutdown – Stops operation automatically once the program completes.
  • Program Error Alarm – Alerts users in case of abnormal operation.

 

Decalcification Time

Mouse Bone Tissue Decalcification Time Table
Bone SiteDecalcification System Time (Days)Traditional Shaker Decalcification Time (Days)
Femur2-37-10
Tibia2-37-10
Ulna & Radius2-37-10
Humerus2-37-10
Elbow Joint3-47-10
Shoulder Joint3-47-10
Knee Joint3-47-10
Ankle Joint4-57-10
Phalanges3-47-10
Nasal Cavity47-10
Skull27-10
Cochlea27-10
Teeth37-10
Spine37-10
Coccyx4-57-10
Note: Both methods use EDTA Decalcification Solution (G1105) with a solution volume ≥20× sample volume. Decalcification is performed at room temperature, with solution replacement every 3 days.

 

Rat Bone Tissue Decalcification Time Table
Bone SiteDecalcification System Time (Days)Traditional Shaker Decalcification Time (Days)
Femur14-1630-35
Tibia14-1630-35
Ulna & Radius14-1630-35
Humerus14-1630-35
Elbow Joint18-2030-35
Shoulder Joint18-2030-35
Knee Joint18-2030-35
Ankle Joint28-3565-70
Phalanges28-3565-70
Nasal Cavity15-1830-35
Skull18-2030-35
Cochlea13-1530-35
Teeth18-2030-35
Spine18-2030-35
Coccyx20-2540-45
Note: Both methods use EDTA Decalcification Solution (G1105) with a solution volume ≥20× sample volume. Decalcification is performed at room temperature, with solution replacement every 3 days.

 

Rabbit Bone Tissue Decalcification Time Table
Bone SiteDecalcification System Time (Days)Traditional Shaker Decalcification Time (Days)
Femur45-6090-95
Tibia45-6090-120
Ulna & Radius45-60110-120
Humerus45-6090-95
Elbow Joint45-6090-95
Shoulder Joint45-6090-95
Nasal Cavity35-4590-95
Skull50-6090-95
Cochlea30-3560-65
Teeth50-6090-95
Spine45-6090-95
Coccyx50-6090-95
Note: Both methods use EDTA Decalcification Solution (G1105) with a solution volume ≥20× sample volume. Decalcification is performed at room temperature, with solution replacement every 3 days.;

 

Important Note: Decalcification time is not absolute and can vary significantly depending on factors such as tissue morphology and bone age. For instance, incomplete removal of skin or muscle, or the presence of other materials in the tissue may affect the decalcification period. For larger specimens, pre-sectioning is recommended to shorten processing time and improve decalcification efficiency.

Technical Parameters

Cat.No.SDH-5
Dimensions520×446×423mm
Weight60Kg
Temperature Range4~60
Timer Range1 min~99 days 24 hrs 59 min
Heating Time (RT→60°C),≤2 hrs
Reaction Chamber Capacity5L
Module Temperature Uniformity≤±1

 

Structural  Diagram

 

 